Left Side Panel

Thermo Scientific
To replace the left side panel of the ISQ 7000 instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
Note You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to loosen the left side panel screw from the interior front
panel.
3. Close the front door.
4. Slide the left side panel toward the back of the instrument and lift it off.
Figure 291. Replacing the Left Side Panel
5. Replace the left side panel and tighten the screw holding it in place.
6. Restart your system by following the instructions in
3
Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
"Shutting Down the ISQ 7000 System"
"Restarting the System"
ISQ 7000 Hardware Manual
on
page
165.