Advanced Maintenance
Replacing the Boards in the ISQ 7000 Mass Spectrometer
302
ISQ 7000 Hardware Manual
4. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
Figure 321. Disconnecting the Electron Multiplier Plate
Electron Multiplier
Plate Screws
5. Lift the electron multiplier and plate out of the instrument and set them aside. You
should now be able to view the 4-pin feedthrough from the top of the instrument.
6. Disconnect the four wires from the 4-pin feedthrough.
Note Each wire must be reattached to the same connector later in this process, so make a
note of their placement.
Figure 322. Disconnecting the Wires
4-pin
Connector
Wires