Source To Ion Guide Spacer - Thermo Scientific ISQ 7000 Hardware Manual

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Ion Source

Source to Ion Guide Spacer

212
ISQ 7000 Hardware Manual
To replace the source to ion guide spacer
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Use a T10 Torxhead screwdriver to disconnect the lens heater block grounding strap from
the lens heater block, but leave it attached to the tray
Figure 200. Disconnecting the Lens Heater Block Grounding Strap
Lens Heater Block
Lens Heater Block
Grounding Strap
page
151.
Screw
Thermo Scientific