Advanced Maintenance
Replacing Components of the Ion Source

Lens Heater Block Grounding Strap

210
ISQ 7000 Hardware Manual
To replace the lens heater block grounding strap
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
3. Use a T10 Torxhead screwdriver to disconnect the screws holding the lens heater block
grounding strap to the lens heater block and analyzer tray.
Figure 198. Replacing the Lens Heater Block Grounding Strap
Lens Heater
Block
Lens Heater Block
Grounding Strap
4. Attach a new lens heater block grounding strap and reattach the screws.
5. Once the components are correctly attached to the analyzer tray, follow the instructions
in
"Replacing the Analyzer Tray"
6. Restart your system by following the instructions in
"Shutting Down the ISQ 7000 System"
on
page
162.
"Restarting the System"
on
page
151.
Screws
on
page
165.