6. Pull the dynode power supply straight up and away from the distribution board on the
bottom of the chassis.
7. Replace the dynode/multiplier power supply and install it by aligning its connector with
the connector on the distribution board beneath it. It will drop into the slot when it is in
the right position.
Figure 392. Replacing the Dynode/Multiplier Power Supply
Connector
8. Push the power supply down into the connector.
9. Reattach the screws holding the dynode and multiplier power supply in place.
Figure 393. Reinstalling the Dynode/Multiplier Power Supply
Back
Screws
Dynode/Multiplier
Power Supply
Bottom
Screw
3
Advanced Maintenance
Replacing the Power Supplies
ISQ 7000 Hardware Manual
351