Thermo Scientific
4. Use a T20 Torxhead screwdriver to remove the two screws on the back of the ion source.
Figure 164. Disconnecting the Ion Source from the Analyzer Tray
Screws
5. Slide the source interface board and ion source out from under the tray without damaging
or disconnecting the wires.
Figure 165. Removing the Ion Source from the Analyzer Tray
6. Set the source interface board on its side.
Wires
3
Advanced Maintenance
Replacing Components of the Ion Source
ISQ 7000 Hardware Manual
187