Thermo Scientific ISQ 7000 Hardware Manual page 256

Mass spectrometers
Table of Contents

Advertisement

3
Advanced Maintenance
Replacing Components of the Vacuum Interlock
228
ISQ 7000 Hardware Manual
15. Flip the block over in your hand so that sealing ball and outer ball seal will fall out. Set
them aside or replace them, if necessary.
CAUTION INSTRUMENT DAMAGE: The sealing ball must not be scratched, so be very
careful when handling it and keep it on a clean, lint-free surface.
ATTENTION DOMMAGES À L'INSTRUMENT : la bille d'étanchéité ne doit pas être rayée,
soyez donc très prudent lorsque vous la manipulez et maintenez-la sur une surface propre et
non pelucheuse.
Figure 227. Replacing the Sealing Ball and Outer Ball Seal
Sealing Ball
Outer Ball Seal
16. Pry the thin metal clip out of the groove. It holds all of the other components inside the
valve body.
Figure 228. Replacing the Metal Clip
Clip
Thermo Scientific