Servicing

Optional Manifold Replacement

Optional
Manifold
Replacement
7-48
48iQ Instruction Manual
Use the following procedure to replace the manifold.
Equipment required:
Hex wrench, 9/16
Hex drive, 9/64
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Unplug three electrical connections (J5, J6, and J8) from the Step POL
board 1.
Figure 7–50. Replacing the Manifold pt 1
3. Unfasten three nuts. Remove the nuts, front and back ferrules as shown
from span, zero in, sample back panel
(Figure
7–51).