4. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately one inch from the top of the board directly under the white label.
Figure 243. Removing the Lens Driver Board
5. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument.
6. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 244. Opening the Manifold Door
3
Replacing Components of the Manifold
Lens Driver Board
Screw
Screws