Advanced Maintenance
Replacing Components of the Ion Source
192
ISQ 7000 Hardware Manual
17. Slide the source interface board and ion source onto the tray without damaging or
disconnecting the wires.
Figure 174. Sliding the Ion Source onto the Analyzer Tray
IMPORTANT When assembling the source and optics block, take extra care to ensure that
the blocks are centered properly and that the ion source cartridge provided with the
instrument can easily slide in and out of the assembly.
18. Reattach the two screws to the back of the ion source.
Figure 175. Reconnecting the Ion Source to the Analyzer Tray
Wires
Screws