Thermo Scientific
6. Disconnect the vacuum interlock cable from the connector and disconnect the tubing.
Figure 304. Disconnecting the Vacuum Interlock Cable
7. Pinch together the two springs on the left side the vacuum interlock shield to remove it.
Figure 305. Removing the Vacuum Interlock Shield
3
Advanced Maintenance
Replacing Components of the Front Door
Tubing
Vacuum
Interlock
Cable
Connector
Vacuum
Interlock
Shield
Springs
ISQ 7000 Hardware Manual
289