Refilling the Calibrant Reservoir

Thermo Scientific
To refill the calibrant reservoir
1. Open the front door of the ISQ 7000 instrument.
2. Twist the calibrant reservoir cover counter-clockwise and remove it from the calibration
gas module.
Figure 87. Removing the Calibrant Reservoir Cover
3. Fill a syringe with 100 μL of Perfluorotributylamine/PFTBA/FC 43 and insert it into the
calibrant reservoir.
Figure 88. Removing the Calibrant Reservoir from the Cover
Calibrant
Reservoir
1
Performing Routine Maintenance
Maintaining the Calibration Gas Module
Calibration
Gas Module
Calibrant
Reservoir
Cover
Calibrant
Reservoir
Cover
ISQ 7000 Hardware Manual
89