Thermo Scientific ISQ 7000 Hardware Manual page 319

Mass spectrometers
Table of Contents

Advertisement

Thermo Scientific
11. Use a 5.5 mm nut driver to remove the four nuts holding the latch to the chassis.
Figure 307. Removing the Latch from the Chassis
12. Attach a new latch to the chassis.
13. Reattach the screws to the interior front panel.
14. Move the tubing back into place and reattach the fitting to the calibration gas module.
15. Reconnect the wires to the electrical connector.
16. Hold the spring and safety interlock in place as you attach the vacuum interlock shield
and screws.
17. Reattach the screws around the vacuum interlock knob.
18. Reattach the vacuum interlock knob.
19. Reattach the left side panel and tighten the screw.
20. Restart your system by following the instructions in
3
Advanced Maintenance
Replacing Components of the Front Door
Nuts
"Restarting the System"
on
ISQ 7000 Hardware Manual
page
165.
291