Delivery Tubes

Thermo Scientific
To replace the delivery tubes
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
149.
2. Remove the analyzer tray from the instrument by following the instructions in
"Removing the Analyzer Tray from the ISQ 7000 Instrument"
Note You do not have to remove the ion source cartridge or the column. If you are using
hydrogen as a carrier gas, power off the GC.
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 267. Opening the Manifold Door
3
Replacing Components of the Manifold
"Shutting Down the ISQ 7000 System"
on
page
Screws
ISQ 7000 Hardware Manual
Advanced Maintenance
151.
255