Replacing the Calibration Gas Components
360
ISQ 7000 Hardware Manual
24. Reinstall the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
Figure 406. Reinstalling the Lens Driver Board
Connector
25. Push the board down into the connectors.
26. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
reattach the screw holding the lens driver board in place on the chassis.
Figure 407. Reattaching the Lens Driver Board
Lens Driver Board Screw