Figure 52. Removing the Ion Source Cartridge
CAUTION - BURN HAZARD The ion source cartridge may be hot, so use caution.
ATTENTION RISQUE DE BRÛLURE : soyez prudent, car la cartouche de la source d'ions
peut être chaude.
13. Let the ion source cartridge cool down before removing the components from the source
holder.
14. Disassemble the ion source cartridge by removing the locking ring first, then the repeller
spring, then the nut, insulator, and repeller (which comes out in one piece), ion volume,
lens 1, lens 2, and lens 3/RF lens.
IMPORTANT Always wear clean, lint-free gloves when handling the ion source cartridge.
1
Performing Routine Maintenance
Cleaning the Ion Source Cartridge
ISQ 7000 Hardware Manual
63