Thermo Scientific ISQ 7000 Hardware Manual page 74

Mass spectrometers
Table of Contents

Advertisement

1
Performing Routine Maintenance
Maintaining the Foreline Pump
46
ISQ 7000 Hardware Manual
30. Disconnect the foreline pump power cable from the System Mech Pump connector on
the back of the instrument.
31. Disconnect the foreline vacuum hose from the foreline pump.
32. Place the foreline pump on a bench.
WARNING LIFTING HAZARD: Use a proper lifting technique with the foreline pump
because it weighs approximately 22 kg (50 lbs).
AVERTISSEMENT RISQUE SOULÈVEMENT : utilisez une technique de soulèvement
appropriée de la pompe préliminaire, étant donné qu'elle pèse environ 22 kg (50 lb).
33. Drain the old oil.
WARNING HAND AND CHEMICAL HAZARD: Wear impermeable laboratory gloves when
changing oil.
AVERTISSEMENT RISQUE PHYSIQUE ET CHIMIQUE : portez des gants de laboratoire
imperméables lorsque vous changez l'huile.
a. Remove one of the oil filler plugs.
b. Remove the drain plug and allow the oil to drain into a suitable container.
c. Dispose of the spent oil according to local environmental regulations.
d. Replace the drain plug.
34. Add fresh oil.
WARNING FIRE HAZARD: Use only factory-approved foreline pump oil. Using
unapproved foreline pump oil may result in a fire or explosion.
AVERTISSEMENT RISQUE D'INCENDIE : utilisez uniquement de l'huile pour pompe
préliminaire approuvée par le fabricant, afin d'éviter d'endommager l'instrument ou de
provoquer un incendie.
a. Add oil to the oil filler plug opening halfway between the MIN and MAX level
marks.