E5071C
Dim DReal As Double
SCPI.CALCulate(1).FSIMulator.BALun.DZConversion.BPORt(1).REAL = 200
DReal = SCPI.CALCulate(1).FSIMulator.BALun.DZConversion.BPORt(1).REAL
Related objects
SCPI.CALCulate(Ch).FSIMulator.BALun.DZConversion.BPORt(Bpt).IMAGinar
y
SCPI.CALCulate(Ch).FSIMulator.BALun.DZConversion.BPORt(Bpt).Z0.R
SCPI.CALCulate(Ch).FSIMulator.BALun.DZConversion.STATe
SCPI.CALCulate(Ch).FSIMulator.STATe
Equivalent key
> Fixture Simulator > Diff ZConversion > Port1(bal) Real|Port2(bal)
Analysis
Real|Port3(bal) Real
Equivalent SCPI command
Syntax
:CALCulate{[1]-160}:FSIMulator:BALun:DZConversion
:BPORt{[1]|2}:REAL <numeric>
:CALCulate{[1]-160}:FSIMulator:BALun:DZConversion
:BPORt{[1]|2}:REAL?
Query response
{numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:FSIM:BAL:DZC:BPOR1:REAL 300"
20 OUTPUT 717;":CALC1:FSIM:BAL:DZC:BPOR1:REAL?"
30 ENTER 717;A