Connector
pin side of cable
Transducer
Strain gage, full bridge
six-wire configuration
CX23-R and EX23-R
Pin
Connection
1
Excitation (-)
Zeroing pulse (-)
2
Excitation (+)
Zeroing pulse (+)
3
Voltage input 10 V (+), 60 V (+)
4
Signal ground
jumper to pin 5
5
Ground cable detect
jumper to pin 4
6
Current input 20 mA (+)
7
Measurement signal (+)
Voltage input 100 mV (+)
f
(-)
1
8
Measurement signal (-)
f
(+)
1
9
Active sensor supply 5...24 V (0 V)
10
Active sensor supply 5...24 V (+)
11
Sense (-)
f
(-)
2
12
Sense (+)
f
(+)
2
13
TEDS (-)
Ground frequency measurement
14
TEDS (+)
Shield
Shield
NOTE
Connection between pins 4 and 5 is necessary for all transducers. Note that the
sensor connector must have a connection between pins 1 and 11 for
compatibility with the MX1615B-R module.
5.3.2 MX1615B-R Bridge Module (Strain Gage Amplifier)
Up to 16 individually configurable inputs can be connected to the MX1615B-R,
including strain gage-based transducers, standardized voltage and resistance or
resistance-based measurements (for example, a PT100 resistance thermometer).
a4275-8.0 en HBM: public
5 SomatXR hardware
Wire Color
(1-KAB183 or 1-KAB184)
Black
Blue
White/Black
Red/Black
Pink/Black
Yellow/Black
White
Red
Brown
Yellow
Grey
Green
Grey/Black
Green/Black
--
MX1615B-R
●
Wiring diagram
99