ASA
DMP41
Effect:
Output of device identification.
Example:
IDN?(x)
HBM,DMP41, 4D:5B:B9:02:00:00:1.0.3.2(y)
Company, device designation, serial number, version num
ber of software.
5.3
Amplifier settings
5.3.1
Amplifier input
Amplifier Sensor Adaptation
Input bridge excitation voltage and transducer type
Syntax:
ASA p1,p2(x)
Parameter:
p1
Code number of required bridge excitation voltage
p1
Bridge excitation voltage
1
2.5 V
2
3
10 V
Parameter:
p2
Code number of the required input sensitivity
p2
Input sensitivity
1
2.5 mV/V
2
5 mV/V (excitation 2.5 V or 5 V)
3
10 mV/V (excitation 2.5 V)
Response:
Acknowledge-
ment
0
Command executed
?
A03474_04_E00_00 HBM: public
Command set of the HBM interpreter
5 V
Significance
Error
Amplifier settings
39