GN610, GN611 Isolation and type testing
1.2
Basic versus reinforced insulation
For reference below one can find the basic insulation and supplementary
insulation as well as the reinforced insulation test values for CATII.
Table 1.2: Test voltages for testing electric strength of solid insulation in
measuring circuits of measurement category II (IEC 61010
Nominal
voltage line to
neutral a.c
r.m.s. or d.c. of
MAINS being
measured. (V)
≤ 150
> 150 ≤ 300
> 300 ≤ 600
> 600 ≤ 1000 3310
To protect a user from hazardous voltages there are several means of
protection possible. As one can see below basic insulation + supplementary
insulation is a possibility but also reinforced isolation is a means of protection.
The test voltages are different per means as can be found in the above table.
Test voltage
5 s a.c. test V a.c. r.m.s.
Basic
Reinforced
insulation and
insulation
supplementary
insulation
840
1390
1390
2210
2210
3510
5400
- 2 - 30:2010)
Impulse test V peak
Basic
Reinforced
insulation and
insulation
suplementary
insulation
1550
2500
2500
4000
4000
6400
6000
9600