Balancing
This ensures that the effect on tolerances of zero point, sensitivity, cable
effects and amplifier sensitivity are compensated. This calibration must take
the whole measurement chain into account.
For instance, when the position of the probe tip or plunger corresponds to the
gauge block, the output signal from the displacement transducer (nominal
sensitivity 80mV/V ±1%) has to be assigned to a display or to an output signal
from the amplifier.
5.2.3
Calibration with input of characteristics
The sensitivity of WA transducers using 80mV/V±1% can be input directly
when using the conveniently designed amplifiers in the MGCplus, MVD2555 or
PME series.
Notice
Please ensure that the sensitivity at an excitation voltage of 2.5V
and a car
rms
rier frecuency of 4800 Hz has been ascertained.
WA...