Distribution Board - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument

Distribution Board

350
TSQ 9610 Hardware Manual
The distribution board lies on the bottom of the TSQ 9610 instrument chassis.
To replace the distribution board
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139.
Note You do not have to remove the ion source cartridge or the column. If you are using
hydrogen as a carrier gas, power off the GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 388. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
CAUTION - INSTRUMENT DAMAGE Limit handling of the boards as much as possible.
Do not touch any board components. Hold the board by the sheet metal bracket only.
"Shutting Down the TSQ 9610 System"
Screws
Screw on Front
of Instrument