Front Door Latch

Thermo Scientific
To replace the latch on the front door
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
There are two major parts to front door latch. One is located on the front door, and one is on
the chassis. If you only need to replace the part on the front door, simply open the door and
follow the instructions below.
1. Use a T10 Torxhead screwdriver to remove the two screws holding the latch onto the
front door.
Figure 350. Removing the Latch from the Front Door
Screws
2. Replace the latch and close the front door of the instrument.
3. To replace the latch on the chassis, shut down your system by following the steps in
"Shutting Down the TSQ 9610 System"
4. Slide the top cover panel off the instrument.
5. Open the front door of the GC and disconnect the column and transfer line nut.
6. Move the TSQ 9610 instrument away from the GC so you can access the transfer line on
the right side.
WARNING BURN HAZARD: The transfer line may be hot, so be careful.
3
Replacing Components of the Front Door
Latch
on
page
139.
TSQ 9610 Hardware Manual
Advanced Troubleshooting
315