Front Door Hinges and Supports

Thermo Scientific
To replace the hinges and supports that connect the front door to the chassis
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139. You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 347. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
3. Lift the left hand front panel off the instrument.
3
Replacing Components of the Front Door
"Shutting Down the TSQ 9610 System"
Screws
TSQ 9610 Hardware Manual
Advanced Troubleshooting
Screw on Front
of Instrument
313