Right Side Panel - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Covers of the TSQ 9610 Instrument

Right Side Panel

308
TSQ 9610 Hardware Manual
To replace the right side panel of the TSQ 9610 instrument
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139.
Note You do not have to remove the ion source cartridge. However, you must vent the
instrument and remove the column.
2. Remove the column from the transfer line completely.
3. Slide the top cover panel toward the back of the instrument and lift it off.
4. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
Figure 343. Removing the Right Side Panel Screws
5. Replace the right side panel and tighten the screws.
6. Replace the top cover panel and tighten the screw holding it in place.
7. Reinsert the column to the proper length and tighten the transfer line nut. See
a Column"
on
page 13
8. Restart your system by following the instructions in
Right Side
Panel Screws
for more information.
"Shutting Down the TSQ 9610 System"
"Restarting the System"
"Replacing
on
page
158.
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents