Parameter
Item
numeric_value
Example
DISP:WIND1:ANAL:LOW –50
Reset condition
On reset, the value is set to –70 dBm for both the upper and lower windows.
Query
DISPlay[:WINDow[1]|2]:ANALog:LOWer? [MIN|MAX]
This query returns the current setting of the analog meter lower scale limit, or the
value associated with MIN or MAX. The response format is <NR3>. The units in
which the results are returned are determined by the current settings of
UNIT:POWer and CALCulate:RELative:STATe as shown in
Query example
DISP:WIND1:ANAL:LOW?
Keysight N432A Programming Guide
Description/Defaul t
A numeric value for the analog meter lower scale limit:
– DEF: The default value is –70 dBm
– MIN: –150 dBm
– MAX: 230 dBm
The units used are determined by the current settings of
UNIT:POWer and CALCulate:RELative:STATe as
shown in
Table
6-1.
This command sets the analog meter lower
scale limit of the upper window to –50 dBm.
This command queries the lower scale limit set on
the analog meter in the upper window.
DISPlay Subsystem
Range of values
–150 to 230 dBm
DEF
MIN
MAX
Table
6-1.
6
261