16
TRIGger Subsystem
TRIGger[:SEQuence[1]]:IMMediate
This command provides a one time override of the normal process of the
downward path through the wait-for-trigger state. It causes the immediate exit of
the event detection layer if the trigger system is in this layer when the command is
received. In other words, the N432A stops waiting for a trigger and takes a
measurement ignoring any delay set by TRIG:DELay.
Syntax
Example
TRIG:IMM
490
TRIG
:SEQ
This command initiates a measurement on the N432A.
1
:IMM
Keysight N432A Programming Guide