MEMory[:METer][1]:CATalog:STATe?
This query is used to list the save/recall registers.
The N432A returns the data in the form of two numeric parameters and as many
strings as there are save/recall registers.
<numeric_value>,<numeric_value>{,<string>}
– The first numeric parameter indicates the amount of memory (in bytes) used
– The second numeric parameter indicates the memory (in bytes) available for
– Each string parameter returned indicates the name, type, and size of a save/
A sample response may look like the following:
0,16190,"State1,STAT,0","State2,STAT,0" .........
Syntax
Example
MEM:CAT:STAT?
Keysight N432A Programming Guide
for the storage of registers.
the storage of registers.
recall register:
<name>,<type>,<size>
– <name> indicates the name of the save/recall register.
– <type> indicates STAT for the save/recall register.
– <size> indicates the size of the save/recall register in bytes.