Keysight N432A Programming Manual page 67

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

Using INITiate and FETCh?
*RST
CONF1
INIT1?
FETC1?
*RST
CONF2
INIT1?
FETC2?
Example 2 – Specifying the source list parameter
The CONFigure command and READ? query have three optional parameters, an
expected power value, a resolution, and a source list. These parameters must be
entered in the specified order. Parameters may be defaulted from the right by
omitting them, or anywhere by substituting the keyword DEFault. DEFault is
used as a placeholder.
The following examples use the source list parameter to specify the measurement
channel as channel A. The expected power and resolution parameters are set to
their default values, leaving them at their current settings. The measurement is
carried out on the upper window.
Although the READ? and FETCh? queries have three optional parameters, it is not
necessary to define them as shown in these examples. If they are defined, they
must be identical to those defined in the CONFigure command, otherwise an error
will occur.
Using READ?
ABOR1
CONF1 DEF,DEF,(@1)
READ1?
Keysight N432A Programming Guide
Resets the N432A.
Configures the upper-window measurement – sets to a single measurement by default.
Sets the channel in a wait-for-trigger state.
Triggers the upper-window measurement, then retrieves the measurement reading.
Resets the N432A.
Configures the lower-window measurement – sets to a single measurement by default.
Sets the channel in a wait-for-trigger state.
Triggers the lower-window measurement, then retrieves the measurement reading.
Aborts the measurement.
Configures the upper window to make a measurement using the
current expected power and resolution settings.
Initiates and retrieves the upper-window measurement.
N432A Remote Operation