Keysight N432A Programming Manual page 70

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

1
N432A Remote Operation
Example 4 – Specifying the resolution parameter
The previous examples detailed the use of the expected value and source list
parameters. The resolution parameter is used to set the resolution of the specified
window.
The following example uses the resolution parameter to specify a resolution
setting of 3. This setting represents three significant digits if the measurement
suffix is W or %, and 0.01 dB if the suffix is dB or dBm (for further details on the
resolution parameter, refer to the commands in
Commands). Also, in this example, the expected power level and source list
parameters are set to their default values. The expected power level is left
unchanged at its current setting. As the source list parameter is the last specified
parameter, you do not have to specify DEF.
Using READ?
ABOR1
CONF1 DEF,3
READ1?
Some finetuning of the above program segment can be carried out, for example,
by turning the trigger delay off, as shown below.
1 ABOR1
2 CONF1 DEF,3
3 TRIG1:DEL:AUTO OFF
4 READ1?
70
Aborts the measurement.
Configures the upper window to make a measurement using the current settings
of the expected power level and source list, and a resolution setting of 3.
Initiates and retrieves the upper-window measurement.
Chapter
3,
MEASurement
Keysight N432A Programming Guide