Keysight N432A Programming Manual page 359

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

Query
[SENSe[1]:]CORRection:CSET[1]:[SELect]?
This query returns the name of the selected sensor calibration table as a quoted
string. If no table is selected, an empty string is returned.
Query example
CORR:CSET?
Error messages
– If the value entered for <string> is not valid, error –224,"Illegal parameter
– If a table with the name specified by <string> does not exist, error –256,"File
– When a sensor calibration table is selected, the N432A verifies that the
Keysight N432A Programming Guide
Queries the sensor calibration table currently used.
value" occurs.
name not found" occurs.
number of calibration points defined is one more than the number of frequency
points defined. When a frequency-dependent offset table is selected, the
N432A verifies that the number of offset points defined is equal to the number
of frequency points defined. If this is not the case, error –226,"Lists not the
same length" occurs.