Keysight N432A Programming Manual page 351

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

[SENSe[1]:]RVALue <numeric_value>
This command sets the resistance value for Ruser. Both the Rmeas and Ruser
values correspond to the current bridge resistance setting. The value specified for
Ruser must be within a ±10% tolerance level of the bridge resistance. If this value
exceeds the tolerance level, it will be clipped to the upper/lower limit within
tolerance. To measure the value for Ruser, refer to the N432A Thermistor Power
Meter User's Guide.
Syntax
Parameter
Item
numeric_value
Reset condition
On reset, the resistance value of Ruser is not affected.
Example
RVAL 110
Keysight N432A Programming Guide
:
SENS
1
Description/Defaul t
A numeric value for the Ruser resistance setting
This command sets the resistance value for Ruser to 110
Space
RVAL
?
SENSe Subsystem
numeric_value
Range of values
Within a ±10% tolerance level of
the current bridge resistance
setting
12
Ω
.
351