16
TRIGger Subsystem
INITiate:CONTinuous:ALL <boolean>
This command sets all trigger sequences to be continuously initiated.
If INITiate:CONTinuous:ALL is set to:
– ON, trigger sequences are set to be continuously initiated.
– OFF, trigger sequences are not set to be continuously initiated.
Syntax
Example
INIT:CONT:ALL ON
Reset condition
On reset (*RST), this command is set to OFF.
On preset (SYSTem:PRESet) and instrument power up, INITiate:CONTinuous:ALL
is set to ON.
476
INIT
:CONT
Space
:ALL
This command sets all trigger sequences to be
continuously initiated.
Keysight N432A Programming Guide
0|OFF
1|ON
?