Keysight N432A Programming Manual page 347

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

[SENSe[1]:]BRESistance <character_data>
This command sets the value of the bridge resistance for the thermistor sensor.
Syntax
Parameter
Item
character_data
Reset condition
On reset, the bridge resistance is not affected.
Example
BRES R100
Keysight N432A Programming Guide
:
SENS
1
Description/Defaul t
Character data representing a bridge resistance value
where:
– R100 indicates 100 Ω
– R200 indicates 200 Ω
– R300 indicates 300 Ω
– R400 indicates 400 Ω
This command sets the bridge resistance to 100
SENSe Subsystem
Space
BRES
?
12
character_data
Range of values
R100
R200
R300
R400
Ω
.