Parameters; Example - Keysight N432A Programming Manual

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

3
MEASurement Commands

Parameters

Refer to
parameters in this command.
Item
expected_value
resolution
source list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command
[b] When the measurement result is linear, this parameter represents the number of significant digits. When the

Example

CONF DEF,2,(@1)
154
"Optional parameters"
Description/Defaul t
A numeric value for the expected power level. The units
of measurement are dBm and W. The default unit is
defined by UNIT:POWer.
A numeric value for the resolution. If unspecified, the
current resolution setting is used.
The channel which the command is implemented on. If
unspecified, the current window setup is used.
subsystems. The parameters must be entered in the specified order. If parameters are omitted, they default from
the right. The parameter DEFault is used as a placeholder. Specifying DEF leaves the parameter value unchanged.
measurement result is logarithmic, 1 to 4 represent 1, 0.1, 0.01, and 0.001 respectively.
on page 146 for additional details on the
This command configures the upper window/upper
measurement to measure power using the current
expected power level and a resolution setting of 2.
Keysight N432A Programming Guide
Range of values
Sensor dependent
[a]
DEF
[b]
1 to 4
1.0, 0.1, 0.01, 0.001
[a]
DEF
(@1)