SERVice Subsystem
SERVice:CALibrator:ADJust:FINE <numeric_value>
This command adjusts the 1 mW reference calibrator output in the fine scale.
Syntax
Parameter
Item
numeric_value
Example
SERV:CAL:ADJ:FINE 500
Reset condition
On reset, the power reference level is not affected.
Query
SERVice:CALibrator:ADJust:FINE?
This query returns the current power reference level in the unsigned integer 16
format.
390
SERV
:CAL
Description/Defaul t
Adjusts the 1 mW power reference level increment
fine by 1
:ADJ
:FINE
This command sets the power reference level
to 500.
numeric_value
Space
DEF
?
Range of values
0 to 1023 (Unsigned Integer
16 format)
Keysight N432A Programming Guide