STATus:OPERation:ULFail[:SUMMary]
The operation status upper limit fail summary register contains information on the
upper limit fail status of the N432A.
The following bits in the register are used by the N432A:
Bit
0 to 2
3
4
5
6
7 to 15
The appropriate bits are set if a window upper limit test fails.
These bits are cleared if a measurement is made and the test is enabled and
passes.
Syntax
Keysight N432A Programming Guide
Weight/Decimal value
–
8
16
32
64
–
STAT
:OPER
Definition
Not used
Upper window upper measurement ULFail status
Lower window upper measurement ULFail status
Upper window lower measurement ULFail status
Lower window lower measurement ULFail status
Not used
:ULF
:SUMM
STATus Subsystem
14
421