Keysight N432A Programming Manual page 454

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

15
SYSTem Subsystem
Data corrupt or stale
230
This occurs when a FETC? is attempted and either a reset has been received or the N432A state has changed
such that the current measurement is invalidated (for example, a change of frequency setting or triggering
conditions)
Data questionable;Input Overload
231
The power input to the channel exceeds the maximum range of the sensor
Data questionable;Input Overload ChA
231
The power input to Channel A exceeds the maximum range of the sensor
Data questionable;Lower window log error
231
This indicates that a difference measurement in the lower window has given a negative result when the units
of measurement were logarithmic
Data questionable;Upper window log error
231
This indicates that a difference measurement in the upper window has given a negative result when the units
of measurement were logarithmic
Data questionable;ZERO ERROR
231
N432A zeroing failed. The most likely cause is attempting to zero when some power signal is being applied to
the sensor.
Data questionable;ZERO ERROR ChA
231
N432A zeroing failed on Channel A. The most likely cause is attempting to zero when some power signal is
being applied to the sensor.
Data questionable;Measurement Over Range
231
The voltage measurement failed as the measured data has exceeded the maximum range
Data questionable;Measurement Under Range
231
The voltage measurement failed as the measured data is below the minimum range
Hardware missing
241
The N432A is unable to execute the command because either no sensor is connected or it expects a
compatible thermistor sensor and one is not connected
System error
310
This indicates that there is a problem with the N432A operation. This error may occur under certain
conditions, for example, the front panel backlight is not turned on when you set the backlight intensity or
there is no non- volatile RAM on board when you configure the warm start state.
Out of memory
321
The N432A required more memory than was available to run an internal operation