Keysight N432A Programming Manual page 129

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

*RCL <NRf>
The *RCL (ReCaLl) command restores the state of the N432A from the specified
save/recall register. This command is only effective when an instrument setup has
been stored previously in the specified register.
Syntax
Parameter
Type
NRf
Example
*RCL 5
Error message
If the register does not contain a saved state, error –224,"Illegal parameter value"
occurs.
Keysight N432A Programming Guide
Space
*RCL
Description/Defaul t
The register number to recall
This command restores the N432A state from register 5.
IEEE-488 Command Reference
NRf
Range of values
1 to 10
129