SERVice:STATe <boolean>
This command enables or disables the warm start feature. The warm start feature
allows you to retain the N432A current states and settings upon power cycle or in
the event of interrupted power. The default state is ON.
Syntax
Example
SERV:STAT 1
Remark
This parameter is not affected by any reset operation and can only be changed by
direct user access.
Query
SERVice:STATe?
This query returns the current state of the warm start feature.
Query example
SERV:STAT?
Error message
If there is no non-volatile RAM on board, error –310,"System error" occurs.
Keysight N432A Programming Guide
SERV
:STAT
This command enables the warm start feature.
Queries whether warm start is enabled or disabled.