MEMory[:METer][1]:TABLe:GAIN[:MAGNitude]
<numeric_value>{,<numeric_value>}
This command is used to enter calibration factors into the sensor calibration table
or offsets into the frequency-dependent offset table, currently selected using
MEMory[:METer][1]:TABLe:SELect. Any previous calibration factor list, or offset
list is cleared before the new calibration factors/offsets are stored.
A maximum of 81 parameters for sensor calibration tables and 80 parameters for
frequency-dependent offset tables can be sent with this command. For sensor
calibration tables only, the first parameter is the reference calibration factor and
each subsequent parameter is a calibration factor point in the sensor calibration
table.
Entries in the frequency lists correspond to entries in the calibration factor/offset
lists as shown in the following table:
For sensor calibration tables, the number of frequency points must be one less
than the number of calibration factor data points. This is verified when the sensor
calibration table is selected using SENSe:CORRection:CSET1:SELect <string>.
Syntax
MEM
:MET
320
Frequency
Frequency 1
"
Frequency 80
:TABL
1
Calibration factor/offset
Reference calibration factor
(For sensor calibration tables only)
Calibration factor/Offset 1
Calibration factor/Offset 80
:GAIN
:MAGN
Keysight N432A Programming Guide
"
,