Keysight N432A Programming Manual page 238

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

4
CALCulate Subsystem
CALCulate[1]|2|3|4:MATH[:EXPRession] <string>
This command sets the specified window/measurement to a single, difference, or
ratio measurement.
Syntax
Parameter
Item
string
[a] Quotes are mandatory. Either single or double quotes may be used.
[b] The mathematical operation will be performed in the linear scale.
Example
CALC2:MATH "(SENS1/SENS1)"
238
:MATH
CALC
1
2
3
4
Description/Defaul t
A single string value specifying the measurement type.
The default value is SENS1.
Space
:EXPR
Range of values
"
"
"
This command sets the lower window/
upper measurement to make a Channel
A/A ratio measurement.
Keysight N432A Programming Guide
string
?
[a]
"
(SENS1)
[a], [b]
"
(SENS1-SENS1)
[a]
"
(SENS1/SENS1)