12
SENSe Subsystem
Parameter
Item
numeric_value
Example
CORR:DCYC 90PCT
Reset condition
On reset, the duty cycle is 1%.
Query
[SENSe[1]:]CORRection:DCYCle|GAIN3[:INPut][:MAGNitude]?
[MIN|MAX]
This query returns the current setting of the duty cycle or the values associated
with MIN and MAX.
Query example
CORR:GAIN3?
368
Description/Defaul t
A numeric value for the duty cycle
– DEF: Default value is 1%
– MIN: 0.001%
– MAX: 99.999%
The unit is PCT, and is optional
This command sets a duty cycle of 90%.
Queries the current setting of the duty cycle.
Range of values
0.001 to 99.999 PCT
DEF
MIN
MAX