Keysight N432A Programming Manual page 321

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

Parameter
Item
numeric_value
Example
MEM:TABL:SEL "Sensor_1"
MEM:TABL:GAIN 97,99.5,97.4
Query
MEMory[:METer][1]:TABLe:GAIN[:MAGNitude]?
This query returns a list of calibration factor/offset points for the currently
selected table.
Query example
MEM:TABL:GAIN?
Error messages
– If more than 81 calibration factors for sensor calibration tables, or 80 offsets
– If a table is not specified using the MEMory:TABLe:SELect command, the
– If any of the calibration factors/offsets are outside of the allowed range, error
Keysight N432A Programming Guide
Description/Defaul t
A numeric value for the calibration factor/offset. The
unit is PCT.
Queries the calibration factor/offset in the current table.
for frequency-dependent offset tables are in the list, error –108,"Parameter not
allowed" occurs.
data cannot be entered and error –221,"Settings conflict" occurs.
–222,"Data out of range" occurs.
MEMory Subsystem
This command enters a reference
calibration factor of 97% and calibration
factors of 99.5% and 97.4% into the
selected sensor calibration table.
10
Range of values
1.0 to 150.0
321