Keysight N432A Programming Manual page 107

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

Table 1-4
Bit
0
1
2
3
4
5
6
7
The Standard Event register is cleared when you:
– send a
– query the event register using the
The Standard Event enable register is cleared when you:
– cycle the N432A power.
– execute an *ESE 0 command.
Keysight N432A Programming Guide
Bit definitions – Standard Event register
Weight/Decimal
Definition
value
Operation complete
1
All overlapped commands following an *OPC command have completed
2
Not used (always set to 0)
Query error
4
A query error occurred
Device-dependent error
8
A device error occurred
Execution error
16
An execution error occurred
Command error
32
A command syntax error occurred
64
User request
Power on
128
Power has been turned off and on since the last time the event register
was read or cleared
*CLS
(clear status) command.
N432A Remote Operation
*ESR?
(event status register) command.
1
107

Advertisement

Table of Contents
loading

Table of Contents