[SENSe[1]:]AVERage:SDETect <boolean>
This command enables or disables step detection. Step detection shortens the
filter time when the input power changes substantially, for the filter output to get
to its final value. Note that this result appears to settle faster, although true
settling to the final value is unaffected.
– Step detection is automatically disabled when TRIG:DEL:AUTO is set to ON
NOTE
– With certain pulsing signals, step detection may operate on the pulses,
Syntax
Reset condition
On reset, step detection is enabled.
Keysight N432A Programming Guide
and the trigger mode is set to free run. Under these circumstances, the value
of SENS:AVER:SDET is ignored. Note also that SENS:AVER:SDET is not set by
the N432A (that is, SENS:AVER:SDET retains its current setting which may
indicate that step detection is turned ON).
preventing the final average from being completed and making the results
unstable. Under these conditions, step detection should be turned OFF.
:
SENS
1
:SDET
AVER
SENSe Subsystem
Space
0|OFF
1|ON
?
12