ABORt[1]
This command removes the N432A from the wait-for-trigger state and places it in
the idle state. It does not affect any other settings of the trigger system. When the
INITiate command is sent, the trigger system responds as it did before ABORt
was executed.
If ABORt is executed while INITiate:CONTinuous is enabled, the measurement
immediately goes into the wait-for-trigger state.
Syntax
Example
ABOR
Keysight N432A Programming Guide
ABOR
1
This command places the N432A in the idle state.