Keysight N432A Programming Manual page 251

Thermistor power meter
Hide thumbs Also See for N432A:
Table of Contents

Advertisement

CALibration[1]:ZERO:AUTO <boolean>|ONCE
This command causes the N432A to perform its zeroing routine when enabled.
This adjusts the N432A for a zero power reading with no power supplied to the
thermistor sensor.
1|ON is ignored in this command. Only ONCE can be specified to initiate zeroing on
the N432A.
0|OFF is only required for the query response and is ignored in this command.
This command assumes that the sensor is not connected to any power source.
Syntax
Example
CAL:ZERO:AUTO ONCE
Reset condition
On reset, automatic zeroing is disabled.
Query
CALibration[1]:ZERO:AUTO?
This query always returns a value of 0.
Keysight N432A Programming Guide
CAL
1
:AUTO
:ZERO
This command causes the N432A to perform a
zeroing routine.
CALibration Subsystem
Space
0|OFF
1|ON
ONCE
?
5
251