TRIGger Subsystem
TRIGger[:SEQuence[1]]:DELay:AUTO <boolean>
This command is used to determine whether or not there is a settling-time delay
before a measurement is made.
When this command is set to:
– ON, the N432A inserts a settling-time delay before taking the requested
– OFF, no settling-time delay is inserted and the N432A makes the measurement
– ONCE, a settling-time delay is inserted before taking the requested
TRIGger:DELay:AUTO is ignored if TRIGger:IMMediate is set to ON.
Syntax
TRIG
:SEQ
Example
TRIG:DEL:AUTO ON
488
measurement and for subsequent measurements. This settling time allows the
internal digital filter to be updated with new values to produce valid and
accurate measurement results. The trigger with delay command allows
settling time for the internal amplifiers and filters. It does not allow time for
sensor delay.
In cases of large power changes, the delay may not be sufficient for a complete
settling. Accurate readings can be assured by taking two successive
measurements for comparison.
immediately after a trigger is received.
measurement, for one measurement only.
1
:DEL
Space
:AUTO
This command enables a delay on the N432A.
Keysight N432A Programming Guide
0|OFF
1|ON
ONCE
?